Structure projection reconstruction from through-focus series of high-resolution transmission electron microscopy images

A structure projection reconstruction method based on contrast transfer function correction of through-focus series of high-resolution transmission electron microscopy images is presented. In this method, defocus values are determined by evaluating phase similarities of the pixels on the Fourier tra...

Full description

Saved in:
Bibliographic Details
Published in:Ultramicroscopy 2012-04, Vol.115, p.50-60
Main Authors: Wan, Wei, Hovmöller, Sven, Zou, Xiaodong
Format: Article
Language:eng
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A structure projection reconstruction method based on contrast transfer function correction of through-focus series of high-resolution transmission electron microscopy images is presented. In this method, defocus values are determined by evaluating phase similarities of the pixels on the Fourier transforms of the images after correction using trial defocus values. Two-fold astigmatism is also determined, by measuring focus variation along different directions. Each image in the series is corrected for the effects of contrast transfer function and then combined into a structure projection image. The method works for both crystalline and non-crystalline objects. Test results with experimental images are presented. Influences of experimental parameters for imaging and effects of crystal thickness on reconstruction are discussed. ► A method to determine defocus and 2-fold astigmatism in HRTEM images is developed. ► The method compares phase similarity in reciprocal space in a through-focus series. ► The effects of contrast transfer function are corrected for each image. ► A structure projection is reconstructed by combining all the corrected images. ► The method can be used for perfect or defect crystals and even amorphous samples.
ISSN:0304-3991
1879-2723
1879-2723