Loading…

Determination of Manganese in Serum Using GFAAS: Serum Reference Values for Manganese in the Adolescent Girls of the DERVAN Cohort

Current proposed method allows for the determination of manganese in serum sample using aqueous standard calibration technique on Graphite Furnace atomic absorption spectrophotometer (GFAAS) using deuterium background correction. This method involves determination of manganese from digested serum sa...

Full description

Saved in:
Bibliographic Details
Published in:Indian journal of clinical biochemistry 2022-10, Vol.37 (4), p.487-493
Main Authors: Chavan, Sagar, Bhat, Rohit, Nandoskar, Ajit, Bhat, Pallavi, Rokade, Shrawani R., Mohire, Rutuja R., Patil, Suvarna N.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Current proposed method allows for the determination of manganese in serum sample using aqueous standard calibration technique on Graphite Furnace atomic absorption spectrophotometer (GFAAS) using deuterium background correction. This method involves determination of manganese from digested serum samples without the use of matrix modifier. Pyrolysis and atomization temperatures are 1200 °C and 2200 °C respectively. The limit of detection (LoD) and limit of quantitation (LoQ) of the method are 0.0097 ng/ml (0.18 nmol/l) and 0.032 ng/ml (0.58 nmol/l) respectively. Validation of the method was carried out using seronorm trace element level-1 serum standard with excellent agreement between measured value and certified value. Accuracy was demonstrated by the spike and recovery study with analytical recovery between 98.8 and 100.6% in serum. The serum reference value for manganese in adolescent girls of rural Konkan region of India range from 4.7 to 215 nmol/l.
ISSN:0970-1915
0974-0422
DOI:10.1007/s12291-022-01021-3