A prototype Multi-X-ray-source array (MXA) for digital breast tomosynthesis
The design and testing of a prototype Multi-X-ray-source Array (MXA) for digital breast tomosynthesis is reported. The MXA is comprised of an array of tungsten filament cathodes with focus cup grid-controlled modulation and a common rotating anode housed in a single vacuum envelope. Prototypes consi...
Saved in:
Published in: | Physics in medicine & biology 2020-12, Vol.65 (23), p.235033-235033 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The design and testing of a prototype Multi-X-ray-source Array (MXA) for digital breast tomosynthesis is reported. The MXA is comprised of an array of tungsten filament cathodes with focus cup grid-controlled modulation and a common rotating anode housed in a single vacuum envelope. Prototypes consisting of arrays of three-source elements and eleven-source-elements were fabricated and evaluated. The prototype sources demonstrated focal spot sizes of 0.3 mm at 45 kV with 50 mA. Measured x-ray spectra were consistent with the molybdenum anode employed, and the tube output (air kerma) was between 0.6 mGy/100 mAs at 20 kV and 17 mGy/100 mAs at 45 kV with a distance of 100 cm. HVL measurements ranged from 0.5 mm Al at 30 kV to 0.8 mm Al at 45 kV, and x-ray pulse widths were varied from 20 ms to 110 ms at operating frequencies ultimately to be limited by source turn-on/off times of ∼1 ms. Initial results of reconstructed tomographic data are presented. |
---|---|
ISSN: | 0031-9155 1361-6560 |