Robotic CT-guided out-of-plane needle insertion: comparison of angle accuracy with manual insertion in phantom and measurement of distance accuracy in animals
Objectives To evaluate the accuracy of robotic CT-guided out-of-plane needle insertion in phantom and animal experiments. Methods A robotic system (Zerobot), developed at our institution, was used for needle insertion. In the phantom experiment, 12 robotic needle insertions into a phantom at various...
Saved in:
Published in: | European radiology 2020-03, Vol.30 (3), p.1342-1349 |
---|---|
Main Authors: | , , , , , , , , , , , , , , |
Format: | Article |
Language: | eng |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Objectives
To evaluate the accuracy of robotic CT-guided out-of-plane needle insertion in phantom and animal experiments.
Methods
A robotic system (Zerobot), developed at our institution, was used for needle insertion. In the phantom experiment, 12 robotic needle insertions into a phantom at various angles in the
XY
and
YZ
planes were performed, and the same insertions were manually performed freehand, as well as guided by a smartphone application (SmartPuncture). Angle errors were compared between the robotic and smartphone-guided manual insertions using Student’s
t
test. In the animal experiment, 6 robotic out-of-plane needle insertions toward targets of 1.0 mm in diameter placed in the kidneys and hip muscles of swine were performed, each with and without adjustment of needle orientation based on reconstructed CT images during insertion. Distance accuracy was calculated as the distance between the needle tip and the target center.
Results
In the phantom experiment, the mean angle errors of the robotic, freehand manual, and smartphone-guided manual insertions were 0.4°, 7.0°, and 3.7° in the
XY
plane and 0.6°, 6.3°, and 0.6° in the
YZ
plane, respectively. Robotic insertions in the
XY
plane were significantly (
p
< 0.001) more accurate than smartphone-guided insertions. In the animal experiment, the overall mean distance accuracy of robotic insertions with and without adjustment of needle orientation was 2.5 mm and 5.0 mm, respectively.
Conclusion
Robotic CT-guided out-of-plane needle insertions were more accurate than smartphone-guided manual insertions in the phantom and were also accurate in the in vivo procedure, particularly with adjustment during insertion.
Key Points
•
Out-of-plane needle insertions performed using our robot were more accurate than smartphone-guided manual insertions in the phantom experiment and were also accurate in the in vivo procedure.
•
In the phantom experiment, the mean angle errors of the robotic and smartphone-guided manual out-of-plane needle insertions were 0.4° and 3.7° in the XY plane (p < 0.001) and 0.6° and 0.6° in the YZ plane (p = 0.65), respectively.
•
In the animal experiment, the overall mean distance accuracies of the robotic out-of-plane needle insertions with and without adjustments of needle orientation during insertion were 2.5 mm and 5.0 mm, respectively. |
---|---|