Interlaboratory comparison of Pseudomonas aeruginosa phage susceptibility testing
Standardized approaches to phage susceptibility testing (PST) are essential to inform selection of phages for study in patients with bacterial infections. There is no reference standard for assessing bacterial susceptibility to phage. We compared agreement between PST performed at three centers: two...
Saved in:
Published in: | Journal of clinical microbiology 2023-12, Vol.61 (12), p.e0061423-e0061423 |
---|---|
Main Authors: | , , , , , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Standardized approaches to phage susceptibility testing (PST) are essential to inform selection of phages for study in patients with bacterial infections. There is no reference standard for assessing bacterial susceptibility to phage. We compared agreement between PST performed at three centers: two centers using a liquid assay standardized between the sites with the third, a plaque assay. Four
phages: PaWRA01ø11 (EPa11), PaWRA01ø39 (EPa39), PaWRA02ø83 (EPa83), PaWRA02ø87 (EPa87), and a cocktail of all four phages were tested against 145
.
isolates. Comparisons were made within measurements at the two sites performing the liquid assay and between these two sites. Agreement was assessed based on coverage probability (CP
), total deviation index, concordance correlation coefficient (CCC), measurement accuracy, and precision. For the liquid assay, there was satisfactory agreement among triplicate measurements made on different days at site 1, and high agreement based on accuracy and precision between duplicate measurements made on the same run at site 2. There was fair accuracy between measurements of the two sites performing the liquid assay, with CCCs below 0.6 for all phages tested. When compared to the plaque assay (performed once at site 3), there was less agreement between results of the liquid and plaque assays than between the two sites performing the liquid assay. Similar findings to the larger group were noted in the subset of 46
.
isolates from cystic fibrosis. Results of this study suggest that reproducibility of PST methods needs further development. |
---|---|