Loading…
A .beta.-gage method applied to aerosol samples
An instrument developed for the routine measurement of aerosol mass by using the beta -particle attenuation method is described and evaluated. Factors affecting the precision and the accuracy of the measurement are discussed in detail. Results of intercomparison studies between the beta -gage method...
Saved in:
Published in: | Environmental science & technology 1981-06, Vol.15 (6), p.680-686 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | An instrument developed for the routine measurement of aerosol mass by using the beta -particle attenuation method is described and evaluated. Factors affecting the precision and the accuracy of the measurement are discussed in detail. Results of intercomparison studies between the beta -gage method and conventional gravimetric methods are presented. The design of the present instrument is particularly well suited for the automatic mass measurements of membrane filters used in modern dichotomous samplers. The precision of such measurements is shown to be less than plus or minus 5 mu g/cm super(2) for counting intervals of 1 min per sample. |
---|---|
ISSN: | 0013-936X 1520-5851 |
DOI: | 10.1021/es00088a006 |