Quantitative and analytical comparison of isodose distributions for shaped electron fields from ADAC Pinnacle treatment planning system and Monte Carlo simulations
We assess the accuracy of ADAC Pinnacle 3 commercial treatment planning system (TPS) in computation of isodose distributions for shaped electron fields. The assessment is based on comparison of dose profiles generated by TPS and a Monte Carlo model for different beam energies, applicator sizes, and...
Saved in:
Published in: | Applied radiation and isotopes 2010-12, Vol.68 (12), p.2174-2180 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We assess the accuracy of ADAC Pinnacle
3 commercial treatment planning system (TPS) in computation of isodose distributions for shaped electron fields. The assessment is based on comparison of dose profiles generated by TPS and a Monte Carlo model for different beam energies, applicator sizes, and percentages of field blocking. Dose differences of up to 14% are observed at the depth of maximum dose. These discrepancies, often ignored in clinical evaluations, are attributable to inadequate modeling of scatter from applicators and blocks by TPS. |
---|---|
ISSN: | 0969-8043 1872-9800 |
DOI: | 10.1016/j.apradiso.2010.04.026 |