Loading…

Estimation of thin-film platinum thermometer precision in the range 13.8–273.16 K and common mathematical description above 70 K

The temperature-resistance dependence of 10 thin-film platinum sensors was experimentally studied between 5 K and 300 K. A method was proposed for evaluating data errors in both coordinates. The orders of magnitude of the errors for the investigated thermometers were ± (5–8) mK for temperature and ±...

Full description

Saved in:
Bibliographic Details
Published in:Cryogenics (Guildford) 1996, Vol.36 (8), p.599-603
Main Authors: Dimitrov, D.A., Georgiev, J.K., Zahariev, A.L., Bivas, I.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The temperature-resistance dependence of 10 thin-film platinum sensors was experimentally studied between 5 K and 300 K. A method was proposed for evaluating data errors in both coordinates. The orders of magnitude of the errors for the investigated thermometers were ± (5–8) mK for temperature and ± (2–3) mΩ for resistance. Common mathematical descriptions of T(R) and R(T) functions above 70 K and a two-point method for individual calibration are proposed.
ISSN:0011-2275