Loading…
Improving CPR performance using an audible feedback system suitable for incorporation into an automated external defibrillator
Background: It has been shown that a computer-based audible feedback system can improve acquisition and retention of basic life support (BLS) skills. This system is being developed to work in association with an automated external defibrillator (AED). Aim: To determine if such a feedback system is l...
Saved in:
Published in: | Resuscitation 2003-04, Vol.57 (1), p.57-62 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background: It has been shown that a computer-based audible feedback system can improve acquisition and retention of basic life support (BLS) skills. This system is being developed to work in association with an automated external defibrillator (AED).
Aim: To determine if such a feedback system is likely to improve the quality of CPR performed by trained nurses whilst using an AED.
Method: Thirty-six general nurses performed 3 min of BLS on a manikin connected to a laptop computer running an experimental software program. After initial testing they were randomly allocated to control or ‘feedback’ groups. Both groups then performed a further 3 min of BLS, but those in the feedback group received audible corrective instructions from the computer when errors of technique were detected.
Results: The group receiving feedback were significantly better than the control group at performing inflations (
P=0.004) and achieving the correct depth of chest compression (
P |
---|---|
ISSN: | 0300-9572 1873-1570 |
DOI: | 10.1016/S0300-9572(02)00400-8 |