Loading…

ST elevation myocardial infarction, or not – This is the question

We present an artefactual ECG created by a digital ECG-recording machine, caused by deletion of the first 80 ms of the QRS complex immediately following the pacing spike in a patient with complete atrio-ventricular block, biventricular pacing and chronic atrial fibrillation. The artefact was mistake...

Full description

Saved in:
Bibliographic Details
Published in:Journal of electrocardiology 2021-09, Vol.68, p.150-152
Main Authors: Margulescu, Andrei D., Thomas, Dewi E.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We present an artefactual ECG created by a digital ECG-recording machine, caused by deletion of the first 80 ms of the QRS complex immediately following the pacing spike in a patient with complete atrio-ventricular block, biventricular pacing and chronic atrial fibrillation. The artefact was mistaken for inferior ST elevation myocardial infarction and the patient underwent unnecessary urgent coronary angiogram. We are not aware of this particular artefact pattern being previously reported in the literature.
ISSN:0022-0736
1532-8430
DOI:10.1016/j.jelectrocard.2021.08.016