Loading…
DETERMINATION OF DIAGNOSTIC REFERENCE LEVEL IN ROUTINE EXAMINATIONS OF DIGITAL RADIOGRAPHY IN MAZANDARAN PROVINCE
Abstract Introduction The main purpose of this study was to determine the diagnostic reference level (DRL) for routine digital radiography examinations in Mazandaran province. Materials and methods Thirteen digital radiographic examinations at 18 high-patient-load radiography centres were investigat...
Saved in:
Published in: | Radiation protection dosimetry 2020-08, Vol.190 (1), p.31-37 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Abstract
Introduction
The main purpose of this study was to determine the diagnostic reference level (DRL) for routine digital radiography examinations in Mazandaran province.
Materials and methods
Thirteen digital radiographic examinations at 18 high-patient-load radiography centres were investigated. The indirect dosimetry method was performed based on the IAEA report. Average entrance skin dose (ESD) and the third quartile of ESD as the DRL were evaluated from the measurement made by a semiconductor dosemeter.
Results
DRL for the examinations of digital radiography was obtained as: Skull (postero-anterior [PA]): 2.2, skull (lateral [LAT]): 2.4, cervical spine (antero-posterior [AP]): 1.6, cervical spine (LAT): 1.7, thoracic spine (AP): 3.6, thoracic spine (LAT): 9.9, lumbar spine (AP): 5.3, lumbar spine (LAT): 11.8, chest (PA): 1.4, chest (LAT): 2.1, abdomen (AP): 4.3, pelvis (AP): 3.2 and hip (AP): 2.1 mGy.
Conclusion
Although DRL was not higher compared with the international organisations’ levels, it can be reduced by adequate training of radiographers. |
---|---|
ISSN: | 0144-8420 1742-3406 |
DOI: | 10.1093/rpd/ncaa074 |