Loading…
Reliability of automated synovial fluid cell counting with Mindray BC‐6800 body fluid mode
Introduction The enumeration and differentiation of nuclear elements in synovial fluid is a cornerstone for diagnosis and follow‐up of many orthopedic and rheumatologic diseases. In this study, we evaluated the analytical performance of Mindray BC‐6800 BF mode (BC‐6800‐BF) for synovial fluid analysi...
Saved in:
Published in: | International journal of laboratory hematology 2017-06, Vol.39 (3), p.337-346 |
---|---|
Main Authors: | , , , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Introduction
The enumeration and differentiation of nuclear elements in synovial fluid is a cornerstone for diagnosis and follow‐up of many orthopedic and rheumatologic diseases. In this study, we evaluated the analytical performance of Mindray BC‐6800 BF mode (BC‐6800‐BF) for synovial fluid analysis.
Methods
Overall, 78 synovial fluids were collected and analyzed with both BC‐6800‐BF and light microscopy. The study also entailed the assessment of limit of blank (LoB), limit of detection (LoD), limit of quantification (LoQ), carryover and linearity.
Results
The LoB for the parameters total cells and white blood cells was 6 × 106 cells/L, and the LoD and LoQ were instead 15 and 16 × 106 cells/L, respectively. Linearity was excellent and carryover was negligible. The agreement between BC‐6800‐BF and light microscopy was satisfactory for all samples pretreated with hyaluronidase, displaying a bias between −5.9% and 8.2%.
Conclusions
The use of BC‐6800‐BF for synovial fluid analysis enables rapid and accurate assessment, especially for total cell and polymorphonuclear counts. The use of BC‐6800‐BF may therefore allow the replacement of optical analysis, especially in samples pretreated with hyaluronidase, thus allowing its routine use for the screening of synovial specimens. |
---|---|
ISSN: | 1751-5521 1751-553X |
DOI: | 10.1111/ijlh.12632 |