Comparison of the High-Energy Photon Reference Fields of PTB and JAEA

A comparison has been conducted between the 6 MeV and 7 MeV (R-F) high-energy photon reference fields of the Physikalisch-Technische Bundesanstalt (PTB) and the Japan Atomic Energy Agency (JAEA). Both fields are set up according to ISO 4037. The results of both participants are in agreement within t...

Full description

Saved in:
Bibliographic Details
Published in:Radiation protection dosimetry 2017-07, Vol.175 (3), p.330-335
Main Authors: Zutz, H, Kowatari, M, Hupe, O
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A comparison has been conducted between the 6 MeV and 7 MeV (R-F) high-energy photon reference fields of the Physikalisch-Technische Bundesanstalt (PTB) and the Japan Atomic Energy Agency (JAEA). Both fields are set up according to ISO 4037. The results of both participants are in agreement within the combined uncertainties, thus the equivalence of the measurements is demonstrated.
ISSN:0144-8420
1742-3406