Direct trace determination of lead in estuarine water using in situ preconcentration of lead hydride on Ir, Zr and W-coated graphite tubes
Methods for the direct trace determination of lead in sea water involving hydride generation, trapping and the atomization of the hydride generated in a graphite furnace has been developed. Different permanent coated graphite tubes: iridium, tungstate and zirconium coated graphite tubes were investi...
Saved in:
Published in: | Analytica chimica acta 1998-07, Vol.368 (3), p.281-289 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Methods for the direct trace determination of lead in sea water involving hydride generation, trapping and the atomization of the hydride generated in a graphite furnace has been developed. Different permanent coated graphite tubes: iridium, tungstate and zirconium coated graphite tubes were investigated for the in situ preconcentration of lead hydride. The results obtained were compared with the ones achieved by the conventional HG-AAS. Trapping on Ir-coated graphite tubes offers the best analytical performances, with detection limits of 60ngl−1. The accuracy was studied using IAEA/W-4 and SLRS-2 reference materials, and the method was applied to several sea water samples. The number of samples analyzed in an hour was 40. |
---|---|
ISSN: | 0003-2670 1873-4324 |
DOI: | 10.1016/S0003-2670(98)00214-1 |