Loading…
Determination of tannic acid by direct chemiluminescence in a FIA assembly
The determination of tannic acid is performed in a FIA assembly on the basis of the analytical output obtained by oxidation of the acid. The analyte solution was daily prepared in a mixture of quinine as sensitiser and perchloric acid and it was injected into a pure water stream acting as a carrier....
Saved in:
Published in: | Talanta (Oxford) 2002-12, Vol.58 (6), p.1243-1251 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The determination of tannic acid is performed in a FIA assembly on the basis of the analytical output obtained by oxidation of the acid. The analyte solution was daily prepared in a mixture of quinine as sensitiser and perchloric acid and it was injected into a pure water stream acting as a carrier. This solution merges with the mixture potassium permanganate in perchloric medium and the resulting chemiluminescence is monitored. The method was applied over the range 0.5–20 mg l
−1 of tannic acid with a LOD 100 μg l
−1. The reproducibility was 2.1% and the sample throughput 54 h
−1. The influence of foreign substances was studied and the new method is applied to the determination of tannic acid in pharmaceutical and galenic formulations in human urine and surface waters. |
---|---|
ISSN: | 0039-9140 1873-3573 |
DOI: | 10.1016/S0039-9140(02)00410-1 |