Possibilities of increasing the interference immunity of radiothermograph applicator antennas for brain diagnostics

The problems of interference immunity of medical radiothermographs are discussed. The possibility of increasing the interference immunity of radiothermographs by optimizing the design of applicator antennas is considered. The electrical parameters of the optimal design of a slot printed applicator a...

Full description

Saved in:
Bibliographic Details
Published in:Sensors and actuators. A. Physical. 2022-04, Vol.337, p.113439, Article 113439
Main Authors: Leushin, Vitaly Yu, Gudkov, Alexander G., Porokhov, Igor O., Vesnin, Sergey G., Sedankin, Mikhail K., Sidorov, Igor A., Solov’ev, Yuriy V., Agasieva, Svetlana V., Chizhikov, Sergey V.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The problems of interference immunity of medical radiothermographs are discussed. The possibility of increasing the interference immunity of radiothermographs by optimizing the design of applicator antennas is considered. The electrical parameters of the optimal design of a slot printed applicator antenna using a human head-phantom were determined using computer simulation methods. A method for evaluating the interference immunity of an applicator antenna placed on the surface of a biological object simulator is proposed. The calculations carried out confirmed the possibility of ensuring the interference immunity of the applicator antennas by placing the absorbing material on the outer surface of the shielding housing of the antenna. The possibility of conducting functional studies of the human brain using medical radiothermographs equipped with interference-proof applicator antennas, without the use of expensive shielded cameras, has been experimentally confirmed.
ISSN:0924-4247
1873-3069
DOI:10.1016/j.sna.2022.113439