Special Features of Analysis and Control of Concentration Inhomogeneity of Complexly Modified Industrial Materials by the Method of Secondary-Ion Mass Spectrometry

Concentration inhomogeneity in iron-carbon alloys is analyzed and controlled. It is shown that application of secondary-ion mass spectrometers with about 200 resolution with respect to the mass numbers and a comparatively large diameter of the primary ion beam (1 – 2 mm) gives results required for d...

Full description

Saved in:
Bibliographic Details
Published in:Metal science and heat treatment 2018-11, Vol.60 (7-8), p.478-481
Main Authors: Anikin, A. A., Venig, S. B., Zhukov, A. G.
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Concentration inhomogeneity in iron-carbon alloys is analyzed and controlled. It is shown that application of secondary-ion mass spectrometers with about 200 resolution with respect to the mass numbers and a comparatively large diameter of the primary ion beam (1 – 2 mm) gives results required for development and control of processes of structure formation in modified alloys and of changes in their physical and mechanical properties.