Readout models for BaFBr0.85I0.15:Eu image plates
The linearity of the photostimulated luminescence process makes repeated image-plate scanning a viable technique to extract a more dynamic range. In order to obtain a response estimate, two semi-empirical models for the readout fading of an image plate are introduced; they relate the depth distribut...
Saved in:
Published in: | Review of scientific instruments 2018-06, Vol.89 (6), p.063101-063101 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The linearity of the photostimulated luminescence process makes repeated image-plate scanning a viable technique to extract a more dynamic range. In order to obtain a response estimate, two semi-empirical models for the readout fading of an image plate are introduced; they relate the depth distribution of activated photostimulated luminescence centers within an image plate to the recorded signal. Model parameters are estimated from image-plate scan series with BAS-MS image plates and the Typhoon FLA 7000 scanner for the hard x-ray image-plate diagnostic over a collection of experiments providing x-ray energy spectra whose approximate shape is a double exponential. |
---|---|