Effects of laryngoscope handle light source on the light intensity from disposable laryngoscope blades

Summary A bench‐top study was performed to assess the effects of different laryngoscope handles on the light intensity delivered from disposable metal or plastic laryngoscope blades. The light intensity from both the handle light sources themselves and the combined handle and laryngoscope blade sets...

Full description

Saved in:
Bibliographic Details
Published in:Anaesthesia 2014-12, Vol.69 (12), p.1331-1336
Main Authors: Milne, A. D., Brousseau, P. A., Brousseau, C. A.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Summary A bench‐top study was performed to assess the effects of different laryngoscope handles on the light intensity delivered from disposable metal or plastic laryngoscope blades. The light intensity from both the handle light sources themselves and the combined handle and laryngoscope blade sets was measured using a custom‐designed testing system and light meter. Five samples of each disposable blade type were tested and compared with a standard re‐usable stainless steel blade using three different handle/light sources (Vital Signs LED, Heine 2.5 V Xenon and 3.5 V Xenon). The light intensity delivered by the disposable blades ranged from 790 to 3846 lux for the different handle types. Overall, the 3.5 V Heine handle delivered the highest light output (p 
ISSN:0003-2409
1365-2044