The variation of the pressure coefficient of manganin sensors at low temperatures
The pressure coefficient of resistance of manganin sensors was determined from isothermal pressure dependences measured at various temperatures in the range 77-352 K in fully hydrostatic conditions using a He-gas pressure system. The coefficient decreases with decreasing temperature, reaches a minim...
Saved in:
Published in: | Measurement science & technology 1999-05, Vol.10 (5), p.343-347 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The pressure coefficient of resistance of manganin sensors was determined from isothermal pressure dependences measured at various temperatures in the range 77-352 K in fully hydrostatic conditions using a He-gas pressure system. The coefficient decreases with decreasing temperature, reaches a minimum at about 110 K with a deviation of 5.0% from its room temperature value, and then increases. Both below and above 110 K the temperature dependence of the coefficient can be approximately treated as linear with the rate equal to -3.80 x 10\+sp\-4\-sp\ K\+sp\-1\-sp\ and 2.56 x 10\+sp\-4\-sp\ K\+sp\-1\-sp\, respectively. (Original abstract - amended) |
---|---|
ISSN: | 0957-0233 1361-6501 |
DOI: | 10.1088/0957-0233/10/5/001 |