Loading…
Characterization of a deep depletion, back-illuminated charge-coupled device in the x-ray range
In this Note we report on the characterization of a "deep depletion" charge-coupled device (CCD) in the keV range. Measurement of quantum efficiency for different energies from 4.5 until 27 keV is given. A measurement of the capability of the CCD to detect single photons is also reported t...
Saved in:
Published in: | Review of scientific instruments 2005-11, Vol.76 (11), p.116101-116101-3 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | In this Note we report on the characterization of a "deep depletion" charge-coupled device (CCD) in the keV range. Measurement of quantum efficiency for different energies from 4.5 until
27
keV
is given. A measurement of the capability of the CCD to detect single photons is also reported together with a determination of the spatial and spectral resolution. A Monte Carlo simulation was used to give a quantitative comparison with the measured spatial resolution. |
---|---|
ISSN: | 0034-6748 1089-7623 |