Enhanced depolarization contrast in polarization-sensitive optical coherence tomography

We demonstrate the first application of the differential depolarization index (DDI) for depolarization imaging in polarization-sensitive optical coherence tomography (PS-OCT). Unlike the widely used degree of polarization uniformity (DOPU), the DDI is independent of the incident polarization state a...

Full description

Saved in:
Bibliographic Details
Published in:Optics letters 2016-05, Vol.41 (10), p.2350-2353
Main Authors: Ortega-Quijano, Noé, Marvdashti, Tahereh, Ellerbee Bowden, Audrey K
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We demonstrate the first application of the differential depolarization index (DDI) for depolarization imaging in polarization-sensitive optical coherence tomography (PS-OCT). Unlike the widely used degree of polarization uniformity (DOPU), the DDI is independent of the incident polarization state and, therefore, more robust to varying system and sample parameters. Moreover, it can be applied to single-input-polarization-state PS-OCT systems, and it overcomes several limitations of the emerging depolarization index used in multiple-input-polarization-state systems. Our results on tissue phantoms and human skin prove that DDI yields significant depolarization contrast improvements compared to DOPU, which highlights its potential for depolarization imaging in PS-OCT.
ISSN:0146-9592
1539-4794
DOI:10.1364/OL.41.002350