A total inverse planning paradigm: prospective clinical trial evaluating the performance of a novel MR-based 3D-printed head immobilization device
•Generation of a 3D-printed mask (3Dp-mask) from MR imaging to manufacture a reproducible positioning and immobilization.•Prospective trial comparing the performance of the 3Dp-mask and standard thermoplastic mask.•3D-p mask yielded comparable inter-fraction motion, while better patient-reported exp...
Saved in:
Published in: | Clinical and translational radiation oncology 2023-09, Vol.42, p.100663-100663, Article 100663 |
---|---|
Main Authors: | , , , , , , , , , , , , , , , , , |
Format: | Article |
Language: | eng |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | •Generation of a 3D-printed mask (3Dp-mask) from MR imaging to manufacture a reproducible positioning and immobilization.•Prospective trial comparing the performance of the 3Dp-mask and standard thermoplastic mask.•3D-p mask yielded comparable inter-fraction motion, while better patient-reported experience.•The proposed approach unlocks a ‘total inverse planning’ paradigm that may improve CNS radiotherapy workflows and value of care.
Brain radiotherapy (cnsRT) requires reproducible positioning and immobilization, attained through redundant dedicated imaging studies and a bespoke moulding session to create a thermoplastic mask (T-mask). Innovative approaches may improve the value of care. We prospectively deployed and assessed the performance of a patient-specific 3D-printed mask (3Dp-mask), generated solely from MR imaging, to replicate a reproducible positioning and tolerable immobilization for patients undergoing cnsRT.
Patients undergoing LINAC-based cnsRT (primary tumors or resected metastases) were enrolled into two arms: control (T-mask) and investigational (3Dp-mask). For the latter, an in-house designed 3Dp-mask was generated from MR images to recreate the head positioning during MR acquisition and allow coupling with the LINAC tabletop. Differences in inter-fraction motion were compared between both arms. Tolerability was assessed using patient-reported questionnaires at various time points.
Between January 2020 - July 2022, forty patients were enrolled (20 per arm). All participants completed the prescribed cnsRT and study evaluations. Average 3Dp-mask design and printing completion time was 36h:50min (range 12h:56min - 42h:01min). Inter-fraction motion analyses showed three-axis displacements comparable to the acceptable tolerance for the current standard-of-care. No differences in patient-reported tolerability were seen at baseline. During the last week of cnsRT, 3Dp-mask resulted in significantly lower facial and cervical discomfort and patients subjectively reported less pressure and confinement sensation when compared to the T-mask. No adverse events were observed.
The proposed total inverse planning paradigm using a 3D-printed immobilization device is feasible and renders comparable inter-fraction performance while offering a better patient experience, potentially improving cnsRT workflows and its cost-effectiveness. |
---|---|