Cam-driven monochromator for QEXAFS
We have developed a cam-drive for quickly tuning the energy of an X-ray monochromator through an X-ray absorption edge for quick extended X-ray absorption spectroscopy (QEXAFS). The data are collected using a 4-channel, 12-bit multiplexed VME analog to digital converter and a VME angle encoder. The...
Saved in:
Published in: | Radiation physics and chemistry (Oxford, England : 1993) England : 1993), 2006-11, Vol.75 (11), p.1962-1965 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We have developed a cam-drive for quickly tuning the energy of an X-ray monochromator through an X-ray absorption edge for quick extended X-ray absorption spectroscopy (QEXAFS). The data are collected using a 4-channel, 12-bit multiplexed VME analog to digital converter and a VME angle encoder. The VME crate controller runs a real-time operating system. This system is capable of collecting 2 EXAFS-scans in 1
s with an energy stability of better than 1
eV. Additional improvements to increase the speed and the energy stability are under way. |
---|---|
ISSN: | 0969-806X 1879-0895 |
DOI: | 10.1016/j.radphyschem.2005.07.047 |