Aseptic loosening from implant-associated osteolysis in reverse shoulder arthroplasty (RSA) may contribute to premature implant failure. Although articular side polyethylene (PE) damage has been well documented in the literature, no studies to date have investigated backside wear in RSA. The aims of...

Full description

Saved in:
Bibliographic Details
Published in:Journal of shoulder and elbow surgery 2022-03, Vol.31 (3), p.545-552
Main Authors: Nicholson, Allen D., Mathew, Joshua I., Koch, Chelsea N., Kontaxis, Andreas, Wright, Timothy, Taylor, Samuel A., Blaine, Theodore A., Dines, Joshua S., Dines, David M., Fu, Michael C., Warren, Russell F., Gulotta, Lawrence V.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c353t-a3b9b1466c16d914badaa4c879add6fb3d6e36cc2d0aa85ab12121d114f4cf323
cites cdi_FETCH-LOGICAL-c353t-a3b9b1466c16d914badaa4c879add6fb3d6e36cc2d0aa85ab12121d114f4cf323
container_end_page 552
container_issue 3
container_start_page 545
container_title Journal of shoulder and elbow surgery
container_volume 31
creator Nicholson, Allen D.
Mathew, Joshua I.
Koch, Chelsea N.
Kontaxis, Andreas
Wright, Timothy
Taylor, Samuel A.
Blaine, Theodore A.
Dines, Joshua S.
Dines, David M.
Fu, Michael C.
Warren, Russell F.
Gulotta, Lawrence V.
description Aseptic loosening from implant-associated osteolysis in reverse shoulder arthroplasty (RSA) may contribute to premature implant failure. Although articular side polyethylene (PE) damage has been well documented in the literature, no studies to date have investigated backside wear in RSA. The aims of this investigation were to (1) document and compare the damage between the backside and articular surface in explanted RSA components, (2) assess whether certain quadrants have a greater propensity for damage, and (3) report the most common mode(s) of backside PE damage. Twenty-one RSA humeral liners retrieved during revision procedures between 2005 and 2014 were included for analysis. The mean time between implantation and extraction was 16 months (10 days–88 months). Diagnoses at the time of revision included dislocation (10), infection (4), mechanical failure (3), loosening (2), and unknown (2). Liners were examined under light microscopy (×10-30 magnification) and damage on the articular and backside of the liner surface was graded using the modified Hood score. The location and damage modality were compared between the articular side and backside of the implant. Damage was noted on the articular surfaces of all 21 liners and on the backside surface of 20 liners. The total damage in all the quadrants was higher on the articular surface than on the backside of the component, with a mean difference in total quadrant damage scores of 11.74 ± 3.53 (P < .001). There was no difference in damage among the quadrants on the backside (P = .44) or the articular surface (P = .08). The articular side exhibited greater scratching, abrasion, and surface deformation than the backside (P < .001). This short-term retrieval study demonstrated that backside PE damage occurs on the humeral component of RSA implants. There was greater damage to the articular side of the liner but wear to the backside was present in almost all liners. The clinical importance of backside wear in RSA and its overall contribution to PE particulate disease and osteolysis needs further investigation.
doi_str_mv 10.1016/j.jse.2021.09.002
format article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2580696517</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S1058274621007084</els_id><sourcerecordid>2580696517</sourcerecordid><originalsourceid>FETCH-LOGICAL-c353t-a3b9b1466c16d914badaa4c879add6fb3d6e36cc2d0aa85ab12121d114f4cf323</originalsourceid><addsrcrecordid>eNp9kD1PwzAQhi0EoqXwA1hQRpaEs524sTpBxZdUiQVmy7EvakLaBDspyr_HVQsjuuFueN5XuoeQawoJBSru6qT2mDBgNAGZALATMqUZZ7HIAE7DDVkes3kqJuTC-xoAZArsnEx4KqjkGZuSxYM2n76yGHVtM2K_HhvcYvSN2kXVNnK4Q-cx8ut2aCy6SLt-7dqu0b4fL8lZqRuPV8c9Ix9Pj-_Ll3j19vy6vF_Fhme8jzUvZEFTIQwVVtK00Fbr1ORzqa0VZcGtQC6MYRa0zjNdUBbGUpqWqSk54zNye-jtXPs1oO_VpvIGm0ZvsR28YlkOQoqMzgNKD6hxrfcOS9W5aqPdqCiovTNVq-BM7Z0pkCo4C5mbY_1QbND-JX4lBWBxADA8uavQKW8q3Bq0lUPTK9tW_9T_AAYvfSg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2580696517</pqid></control><display><type>article</type><title>Backside polyethylene wear in reverse shoulder arthroplasty</title><source>ScienceDirect Freedom Collection 2022-2024</source><creator>Nicholson, Allen D. ; Mathew, Joshua I. ; Koch, Chelsea N. ; Kontaxis, Andreas ; Wright, Timothy ; Taylor, Samuel A. ; Blaine, Theodore A. ; Dines, Joshua S. ; Dines, David M. ; Fu, Michael C. ; Warren, Russell F. ; Gulotta, Lawrence V.</creator><creatorcontrib>Nicholson, Allen D. ; Mathew, Joshua I. ; Koch, Chelsea N. ; Kontaxis, Andreas ; Wright, Timothy ; Taylor, Samuel A. ; Blaine, Theodore A. ; Dines, Joshua S. ; Dines, David M. ; Fu, Michael C. ; Warren, Russell F. ; Gulotta, Lawrence V.</creatorcontrib><description>Aseptic loosening from implant-associated osteolysis in reverse shoulder arthroplasty (RSA) may contribute to premature implant failure. Although articular side polyethylene (PE) damage has been well documented in the literature, no studies to date have investigated backside wear in RSA. The aims of this investigation were to (1) document and compare the damage between the backside and articular surface in explanted RSA components, (2) assess whether certain quadrants have a greater propensity for damage, and (3) report the most common mode(s) of backside PE damage. Twenty-one RSA humeral liners retrieved during revision procedures between 2005 and 2014 were included for analysis. The mean time between implantation and extraction was 16 months (10 days–88 months). Diagnoses at the time of revision included dislocation (10), infection (4), mechanical failure (3), loosening (2), and unknown (2). Liners were examined under light microscopy (×10-30 magnification) and damage on the articular and backside of the liner surface was graded using the modified Hood score. The location and damage modality were compared between the articular side and backside of the implant. Damage was noted on the articular surfaces of all 21 liners and on the backside surface of 20 liners. The total damage in all the quadrants was higher on the articular surface than on the backside of the component, with a mean difference in total quadrant damage scores of 11.74 ± 3.53 (P &lt; .001). There was no difference in damage among the quadrants on the backside (P = .44) or the articular surface (P = .08). The articular side exhibited greater scratching, abrasion, and surface deformation than the backside (P &lt; .001). This short-term retrieval study demonstrated that backside PE damage occurs on the humeral component of RSA implants. There was greater damage to the articular side of the liner but wear to the backside was present in almost all liners. The clinical importance of backside wear in RSA and its overall contribution to PE particulate disease and osteolysis needs further investigation.</description><identifier>ISSN: 1058-2746</identifier><identifier>EISSN: 1532-6500</identifier><identifier>DOI: 10.1016/j.jse.2021.09.002</identifier><identifier>PMID: 34619352</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>Arthroplasty, Replacement, Shoulder - adverse effects ; backside polyethylene wear ; Humans ; implant retrieval ; Osteolysis ; Polyethylene ; polyethylene wear ; Prosthesis Design ; Prosthesis Failure ; Reverse shoulder arthroplasty</subject><ispartof>Journal of shoulder and elbow surgery, 2022-03, Vol.31 (3), p.545-552</ispartof><rights>2021</rights><rights>Copyright © 2021. Published by Elsevier Inc.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c353t-a3b9b1466c16d914badaa4c879add6fb3d6e36cc2d0aa85ab12121d114f4cf323</citedby><cites>FETCH-LOGICAL-c353t-a3b9b1466c16d914badaa4c879add6fb3d6e36cc2d0aa85ab12121d114f4cf323</cites><orcidid>0000-0001-5743-2670 ; 0000-0002-4094-4781 ; 0000-0001-6420-5142 ; 0000-0003-2178-9277 ; 0000-0003-0605-5312 ; 0000-0002-0383-0558</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>315,786,790,27957,27958</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/34619352$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Nicholson, Allen D.</creatorcontrib><creatorcontrib>Mathew, Joshua I.</creatorcontrib><creatorcontrib>Koch, Chelsea N.</creatorcontrib><creatorcontrib>Kontaxis, Andreas</creatorcontrib><creatorcontrib>Wright, Timothy</creatorcontrib><creatorcontrib>Taylor, Samuel A.</creatorcontrib><creatorcontrib>Blaine, Theodore A.</creatorcontrib><creatorcontrib>Dines, Joshua S.</creatorcontrib><creatorcontrib>Dines, David M.</creatorcontrib><creatorcontrib>Fu, Michael C.</creatorcontrib><creatorcontrib>Warren, Russell F.</creatorcontrib><creatorcontrib>Gulotta, Lawrence V.</creatorcontrib><title>Backside polyethylene wear in reverse shoulder arthroplasty</title><title>Journal of shoulder and elbow surgery</title><addtitle>J Shoulder Elbow Surg</addtitle><description>Aseptic loosening from implant-associated osteolysis in reverse shoulder arthroplasty (RSA) may contribute to premature implant failure. Although articular side polyethylene (PE) damage has been well documented in the literature, no studies to date have investigated backside wear in RSA. The aims of this investigation were to (1) document and compare the damage between the backside and articular surface in explanted RSA components, (2) assess whether certain quadrants have a greater propensity for damage, and (3) report the most common mode(s) of backside PE damage. Twenty-one RSA humeral liners retrieved during revision procedures between 2005 and 2014 were included for analysis. The mean time between implantation and extraction was 16 months (10 days–88 months). Diagnoses at the time of revision included dislocation (10), infection (4), mechanical failure (3), loosening (2), and unknown (2). Liners were examined under light microscopy (×10-30 magnification) and damage on the articular and backside of the liner surface was graded using the modified Hood score. The location and damage modality were compared between the articular side and backside of the implant. Damage was noted on the articular surfaces of all 21 liners and on the backside surface of 20 liners. The total damage in all the quadrants was higher on the articular surface than on the backside of the component, with a mean difference in total quadrant damage scores of 11.74 ± 3.53 (P &lt; .001). There was no difference in damage among the quadrants on the backside (P = .44) or the articular surface (P = .08). The articular side exhibited greater scratching, abrasion, and surface deformation than the backside (P &lt; .001). This short-term retrieval study demonstrated that backside PE damage occurs on the humeral component of RSA implants. There was greater damage to the articular side of the liner but wear to the backside was present in almost all liners. The clinical importance of backside wear in RSA and its overall contribution to PE particulate disease and osteolysis needs further investigation.</description><subject>Arthroplasty, Replacement, Shoulder - adverse effects</subject><subject>backside polyethylene wear</subject><subject>Humans</subject><subject>implant retrieval</subject><subject>Osteolysis</subject><subject>Polyethylene</subject><subject>polyethylene wear</subject><subject>Prosthesis Design</subject><subject>Prosthesis Failure</subject><subject>Reverse shoulder arthroplasty</subject><issn>1058-2746</issn><issn>1532-6500</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2022</creationdate><recordtype>article</recordtype><recordid>eNp9kD1PwzAQhi0EoqXwA1hQRpaEs524sTpBxZdUiQVmy7EvakLaBDspyr_HVQsjuuFueN5XuoeQawoJBSru6qT2mDBgNAGZALATMqUZZ7HIAE7DDVkes3kqJuTC-xoAZArsnEx4KqjkGZuSxYM2n76yGHVtM2K_HhvcYvSN2kXVNnK4Q-cx8ut2aCy6SLt-7dqu0b4fL8lZqRuPV8c9Ix9Pj-_Ll3j19vy6vF_Fhme8jzUvZEFTIQwVVtK00Fbr1ORzqa0VZcGtQC6MYRa0zjNdUBbGUpqWqSk54zNye-jtXPs1oO_VpvIGm0ZvsR28YlkOQoqMzgNKD6hxrfcOS9W5aqPdqCiovTNVq-BM7Z0pkCo4C5mbY_1QbND-JX4lBWBxADA8uavQKW8q3Bq0lUPTK9tW_9T_AAYvfSg</recordid><startdate>202203</startdate><enddate>202203</enddate><creator>Nicholson, Allen D.</creator><creator>Mathew, Joshua I.</creator><creator>Koch, Chelsea N.</creator><creator>Kontaxis, Andreas</creator><creator>Wright, Timothy</creator><creator>Taylor, Samuel A.</creator><creator>Blaine, Theodore A.</creator><creator>Dines, Joshua S.</creator><creator>Dines, David M.</creator><creator>Fu, Michael C.</creator><creator>Warren, Russell F.</creator><creator>Gulotta, Lawrence V.</creator><general>Elsevier Inc</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0001-5743-2670</orcidid><orcidid>https://orcid.org/0000-0002-4094-4781</orcidid><orcidid>https://orcid.org/0000-0001-6420-5142</orcidid><orcidid>https://orcid.org/0000-0003-2178-9277</orcidid><orcidid>https://orcid.org/0000-0003-0605-5312</orcidid><orcidid>https://orcid.org/0000-0002-0383-0558</orcidid></search><sort><creationdate>202203</creationdate><title>Backside polyethylene wear in reverse shoulder arthroplasty</title><author>Nicholson, Allen D. ; Mathew, Joshua I. ; Koch, Chelsea N. ; Kontaxis, Andreas ; Wright, Timothy ; Taylor, Samuel A. ; Blaine, Theodore A. ; Dines, Joshua S. ; Dines, David M. ; Fu, Michael C. ; Warren, Russell F. ; Gulotta, Lawrence V.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c353t-a3b9b1466c16d914badaa4c879add6fb3d6e36cc2d0aa85ab12121d114f4cf323</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2022</creationdate><topic>Arthroplasty, Replacement, Shoulder - adverse effects</topic><topic>backside polyethylene wear</topic><topic>Humans</topic><topic>implant retrieval</topic><topic>Osteolysis</topic><topic>Polyethylene</topic><topic>polyethylene wear</topic><topic>Prosthesis Design</topic><topic>Prosthesis Failure</topic><topic>Reverse shoulder arthroplasty</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Nicholson, Allen D.</creatorcontrib><creatorcontrib>Mathew, Joshua I.</creatorcontrib><creatorcontrib>Koch, Chelsea N.</creatorcontrib><creatorcontrib>Kontaxis, Andreas</creatorcontrib><creatorcontrib>Wright, Timothy</creatorcontrib><creatorcontrib>Taylor, Samuel A.</creatorcontrib><creatorcontrib>Blaine, Theodore A.</creatorcontrib><creatorcontrib>Dines, Joshua S.</creatorcontrib><creatorcontrib>Dines, David M.</creatorcontrib><creatorcontrib>Fu, Michael C.</creatorcontrib><creatorcontrib>Warren, Russell F.</creatorcontrib><creatorcontrib>Gulotta, Lawrence V.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Journal of shoulder and elbow surgery</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Nicholson, Allen D.</au><au>Mathew, Joshua I.</au><au>Koch, Chelsea N.</au><au>Kontaxis, Andreas</au><au>Wright, Timothy</au><au>Taylor, Samuel A.</au><au>Blaine, Theodore A.</au><au>Dines, Joshua S.</au><au>Dines, David M.</au><au>Fu, Michael C.</au><au>Warren, Russell F.</au><au>Gulotta, Lawrence V.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Backside polyethylene wear in reverse shoulder arthroplasty</atitle><jtitle>Journal of shoulder and elbow surgery</jtitle><addtitle>J Shoulder Elbow Surg</addtitle><date>2022-03</date><risdate>2022</risdate><volume>31</volume><issue>3</issue><spage>545</spage><epage>552</epage><pages>545-552</pages><issn>1058-2746</issn><eissn>1532-6500</eissn><notes>ObjectType-Article-1</notes><notes>SourceType-Scholarly Journals-1</notes><notes>ObjectType-Feature-2</notes><notes>content type line 23</notes><abstract>Aseptic loosening from implant-associated osteolysis in reverse shoulder arthroplasty (RSA) may contribute to premature implant failure. Although articular side polyethylene (PE) damage has been well documented in the literature, no studies to date have investigated backside wear in RSA. The aims of this investigation were to (1) document and compare the damage between the backside and articular surface in explanted RSA components, (2) assess whether certain quadrants have a greater propensity for damage, and (3) report the most common mode(s) of backside PE damage. Twenty-one RSA humeral liners retrieved during revision procedures between 2005 and 2014 were included for analysis. The mean time between implantation and extraction was 16 months (10 days–88 months). Diagnoses at the time of revision included dislocation (10), infection (4), mechanical failure (3), loosening (2), and unknown (2). Liners were examined under light microscopy (×10-30 magnification) and damage on the articular and backside of the liner surface was graded using the modified Hood score. The location and damage modality were compared between the articular side and backside of the implant. Damage was noted on the articular surfaces of all 21 liners and on the backside surface of 20 liners. The total damage in all the quadrants was higher on the articular surface than on the backside of the component, with a mean difference in total quadrant damage scores of 11.74 ± 3.53 (P &lt; .001). There was no difference in damage among the quadrants on the backside (P = .44) or the articular surface (P = .08). The articular side exhibited greater scratching, abrasion, and surface deformation than the backside (P &lt; .001). This short-term retrieval study demonstrated that backside PE damage occurs on the humeral component of RSA implants. There was greater damage to the articular side of the liner but wear to the backside was present in almost all liners. The clinical importance of backside wear in RSA and its overall contribution to PE particulate disease and osteolysis needs further investigation.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>34619352</pmid><doi>10.1016/j.jse.2021.09.002</doi><tpages>8</tpages><orcidid>https://orcid.org/0000-0001-5743-2670</orcidid><orcidid>https://orcid.org/0000-0002-4094-4781</orcidid><orcidid>https://orcid.org/0000-0001-6420-5142</orcidid><orcidid>https://orcid.org/0000-0003-2178-9277</orcidid><orcidid>https://orcid.org/0000-0003-0605-5312</orcidid><orcidid>https://orcid.org/0000-0002-0383-0558</orcidid></addata></record>
fulltext fulltext
identifier ISSN: 1058-2746
ispartof Journal of shoulder and elbow surgery, 2022-03, Vol.31 (3), p.545-552
issn 1058-2746
1532-6500
language eng
recordid cdi_proquest_miscellaneous_2580696517
source ScienceDirect Freedom Collection 2022-2024
subjects Arthroplasty, Replacement, Shoulder - adverse effects
backside polyethylene wear
Humans
implant retrieval
Osteolysis
Polyethylene
polyethylene wear
Prosthesis Design
Prosthesis Failure
Reverse shoulder arthroplasty
title Backside polyethylene wear in reverse shoulder arthroplasty
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-09-22T14%3A41%3A09IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Backside%20polyethylene%20wear%20in%20reverse%20shoulder%20arthroplasty&rft.jtitle=Journal%20of%20shoulder%20and%20elbow%20surgery&rft.au=Nicholson,%20Allen%20D.&rft.date=2022-03&rft.volume=31&rft.issue=3&rft.spage=545&rft.epage=552&rft.pages=545-552&rft.issn=1058-2746&rft.eissn=1532-6500&rft_id=info:doi/10.1016/j.jse.2021.09.002&rft_dat=%3Cproquest_cross%3E2580696517%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c353t-a3b9b1466c16d914badaa4c879add6fb3d6e36cc2d0aa85ab12121d114f4cf323%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=2580696517&rft_id=info:pmid/34619352&rfr_iscdi=true