Portable Data Acquisition and Fluidic System for Electrochemical Sensors
The recent outbreak of infectious diseases has highlighted the necessity of point-of-care detection compared to central lab analysis for more effective epidemic control. Recent developments in the field of biosensors have allowed sensitive, accurate disease diagnosis using low-cost devices. In this...
Saved in:
Main Authors: | , , , , , , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Request full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The recent outbreak of infectious diseases has highlighted the necessity of point-of-care detection compared to central lab analysis for more effective epidemic control. Recent developments in the field of biosensors have allowed sensitive, accurate disease diagnosis using low-cost devices. In this work, we describe the development of a portable data acquisition and fluidic system for miniature electrochemical biosensors. The data acquisition system was designed as a single printed circuit board and can perform cyclic voltammetry. The fluidic chamber was designed to work with three miniature sensors which are placed on a single platform. Leakage tests were performed to ensure that each chamber allows sensor isolation and avoids any cross- contamination. Measurements using the fabricated potentiostat board were taken and compared with a commercial potentiostat. It was found that the designed potentiostat was able to measure the same resolution and peak separation in cyclic voltammetry measurements. |
---|---|